Accuracy of a cone beam computed tomography-guided surgical stent for orthodontic mini-implant placement.
Author | |
---|---|
Abstract |
:
Abstract Objective: To validate the accuracy of a cone-beam computed tomography (CBCT)-guided surgical stent for orthodontic mini-implant (OMI) placement by quantitatively evaluating the difference between CBCT-prescribed and actual position of mini-implants in preoperative and postoperative CBCT images. Materials and Methods: A surgical stent was fabricated using Teflon-Perfluoroalkoxy, which has appropriate biological x-ray attenuation properties. Polyvinylsiloxane impression material was used to secure the custom-made surgical stent onto swine mandibles. CBCT scanning was done with the stent in place to virtually plan mini-implants using a three-dimensional (3D) software program. An appropriate insertion point was determined using 3D reconstruction data, and the vertical and horizontal angulations were determined using four prescribed angles. A custom-designed surveyor was used to drill a guide hole within the surgical stent as prescribed on the CBCT images for insertion of 32 OMIs. The mandibles with a surgical stent in place were rescanned with CBCT to measure the deviations between the virtual planning data and surgical results. Results: The difference between the prescribed and actual vertical angle was 1.01 ± 7.25, and the horizontal difference was 1.16 ± 6.08. The correlation coefficient confirms that there was no intrarater variability in either the horizontal (R = .97) or vertical (R = .74) vectors. Conclusions: The surgical stent in this study guides mini-implants to the prescribed position as planned in CBCT. Since the statistical difference was not significant, the surgical stent can be considered to be an accurate guide tool for mini-implant placement in clinical use. |
Year of Publication |
:
1969
|
Journal |
:
The Angle orthodontist
|
Date Published |
:
2011 Aug 29
|
ISSN Number |
:
0003-3219
|
DOI |
:
10.2319/060811-374.1
|
Short Title |
:
Accuracy of a cone beam computed tomographyguided surgical stent
|
Download citation |